    A simple and sensitive spectrophotometric flow injection analysis system has been developed for the determination of trace amounts of nitrite. The method is based on the reaction of nitrite with sulfanilamide in acidic medium to form a diazonium salt which then couples with thenoyltrifluoroacetone in alkaline medium to give an azo dye. Beera??s law is obeyed at 470 nm over a nitrite concentration range of 1-1200 I?g L-1, and a detection limit for nitrite is 0.3 I?g L-1 is achieved with a relative standard deviation of 1.4 at 50 I?g L-1 (n=8). An injection rate of about 40 h-1 is possible using this method, which was successfully applied to the determination of nitrite in rain water, tap water, pond water, and river water samples.
    Original languageEnglish
    Pages (from-to)146 - 150
    Number of pages5
    JournalJournal of Flow Injection Analysis
    Volume27
    Issue number2
    Publication statusPublished - 2010