TY - JOUR
T1 - Quantification and significance of fluid shear stress field in biaxial cell stretching device
AU - Thompson, Mark S.
AU - Abercrombie, Stuart R.
AU - Ott, Claus Eric
AU - Bieler, Friederike H.
AU - Duda, Georg N.
AU - Ventikos, Yiannis
N1 - Funding Information:
Acknowledgments This research was supported by the John Fell OUP Research Fund, University of Oxford, by the Hans Jörg Wyss AO Medical Foundation and by the BMBF funded Berlin-Brandenburg Center for Regenerative Therapies. Claus-Eric Ott is supported by the Deutsche Forschungsgemeinschaft (DFG), Contract grant number: SFB 760. The authors are grateful to Dr M. Megahed for allowing the use of the CFD-ACE platform.
PY - 2011/7
Y1 - 2011/7
N2 - A widely used commercially available system for the investigation of mechanosensitivity applies a biaxial strain field to cells cultured on a compliant silicone substrate membrane stretched over a central post. As well as intended substrate strain, this device also provides a fluid flow environment for the cultured cells. In order to interpret the relevance of experiments using this device to the in vivo and clinical situation, it is essential to characterise both substrate and fluid environments. While previous work has detailed the substrate strain, the fluid shear stresses, to which bone cells are known to be sensitive, are unknown. Therefore, a fluid structure interaction computational fluid dynamics model was constructed, incorporating a finite element technique capable of capturing the contact between the post and the silicone substrate membrane, to the underside of which the pump control pressure was applied. Flow verification experiments using 10-μm-diameter fluorescent microspheres were carried out. Fluid shear stress increased approximately linearly with radius along the on-post sub-strate membrane, with peak values located close to the post edge. Changes in stimulation frequency and culture medium viscosity effected proportional changes in the magnitude of the fluid shear stress (peak fluid shear stresses varied in the range 0.09-3.5Pa), with minor effects on temporal and spatial distribution. Good agreement was obtained between predicted and measured radial flow patterns. These results suggest a reinterpretation of previous data obtained using this device to include the potential for a strong role of fluid shear stress in mechanosensitivity.
AB - A widely used commercially available system for the investigation of mechanosensitivity applies a biaxial strain field to cells cultured on a compliant silicone substrate membrane stretched over a central post. As well as intended substrate strain, this device also provides a fluid flow environment for the cultured cells. In order to interpret the relevance of experiments using this device to the in vivo and clinical situation, it is essential to characterise both substrate and fluid environments. While previous work has detailed the substrate strain, the fluid shear stresses, to which bone cells are known to be sensitive, are unknown. Therefore, a fluid structure interaction computational fluid dynamics model was constructed, incorporating a finite element technique capable of capturing the contact between the post and the silicone substrate membrane, to the underside of which the pump control pressure was applied. Flow verification experiments using 10-μm-diameter fluorescent microspheres were carried out. Fluid shear stress increased approximately linearly with radius along the on-post sub-strate membrane, with peak values located close to the post edge. Changes in stimulation frequency and culture medium viscosity effected proportional changes in the magnitude of the fluid shear stress (peak fluid shear stresses varied in the range 0.09-3.5Pa), with minor effects on temporal and spatial distribution. Good agreement was obtained between predicted and measured radial flow patterns. These results suggest a reinterpretation of previous data obtained using this device to include the potential for a strong role of fluid shear stress in mechanosensitivity.
KW - Cell culture
KW - Computational fluid dynamics
KW - Fluid shear stress
KW - Mechanotransduction
UR - https://www.scopus.com/pages/publications/80054062086
U2 - 10.1007/s10237-010-0255-1
DO - 10.1007/s10237-010-0255-1
M3 - Article
C2 - 20853016
AN - SCOPUS:80054062086
SN - 1617-7959
VL - 10
SP - 559
EP - 564
JO - Biomechanics and Modeling in Mechanobiology
JF - Biomechanics and Modeling in Mechanobiology
IS - 4
ER -