Abstract
Induced activity due to photonuclear reactions produced in the vacuum window of the accelerating wave-guide, the x-ray target and the beam-flattening filter of an 18 MV Siemens KDS linac has been studied. Measurements were performed using a high-purity portable germanium detector. Radioisotopes such as 196Au, 57CO, 60Co and other traces were detected one week after the last clinical use of the linac.
Original language | English |
---|---|
Pages (from-to) | N243-N246 |
Number of pages | 4 |
Journal | Physics in Medicine & Biology |
Volume | 49 |
Issue number | 14 |
DOIs | |
Publication status | Published - 21 Jul 2004 |
Externally published | Yes |