T1 - Pacing and implantable cardioverter defibrillator lead perforation as assessed by multiplanar reformatted ECG-gated cardiac computed tomography and clinical correlates
AU - Pang, Benjamin J
AU - Lui, Elaine H
AU - Joshi, Subodh B
AU - Tacey, Mark
AU - Alison, Jeffery F
AU - Seneviratne, Sujith
AU - Cameron, James Donald
AU - Mond, Harry George
PY - 2014
Y1 - 2014
N2 - INTRODUCTION: We aimed to assess the utility of cardiac computed tomography (CT) in the evaluation of right atrial (RA) and right ventricular (RV) pacemaker and implantable cardiac defibrillator lead perforation. METHODS: Images from a 320-slice electrocardiogram-gated cardiac CT scanner were retrospectively independently analyzed by two reviewers for lead position, pericardial effusion, and perforation.Perforation results were correlated with pacing sensing, impedance, and threshold measurements. RESULTS: A total of 52 patients had RV leads and 35 had RA leads. Five of 17 RV apical, one of 35 RV nonapical, and none of the 35 RA leads perforated through the myocardium on CT imaging criteria. Two clinically perforated leads (that had protruded 5 mm and 15 mm from the outer edge of the myocardium)had pericardial effusions and changes in pacing parameters, and required RV lead repositioning. In contrast,there were four apparent radiologic perforations (that had protruded only an average 1.5+/-0.5 mm from the outer edge of the myocardium) that did not require repositioning. These had the radiologic appearance of perforation on cardiac CT; however, they were not associated with pericardial effusions or significant changes in RV pacing lead sensing, impedance, and threshold measurements. CONCLUSIONS: Cardiac CT scanning with multiplanar reformatting is useful for documenting lead position and assessing for possible cardiac perforation. The clinical significance and natural history of leads with only the appearance of perforation on cardiac CT is uncertain.
AB - INTRODUCTION: We aimed to assess the utility of cardiac computed tomography (CT) in the evaluation of right atrial (RA) and right ventricular (RV) pacemaker and implantable cardiac defibrillator lead perforation. METHODS: Images from a 320-slice electrocardiogram-gated cardiac CT scanner were retrospectively independently analyzed by two reviewers for lead position, pericardial effusion, and perforation.Perforation results were correlated with pacing sensing, impedance, and threshold measurements. RESULTS: A total of 52 patients had RV leads and 35 had RA leads. Five of 17 RV apical, one of 35 RV nonapical, and none of the 35 RA leads perforated through the myocardium on CT imaging criteria. Two clinically perforated leads (that had protruded 5 mm and 15 mm from the outer edge of the myocardium)had pericardial effusions and changes in pacing parameters, and required RV lead repositioning. In contrast,there were four apparent radiologic perforations (that had protruded only an average 1.5+/-0.5 mm from the outer edge of the myocardium) that did not require repositioning. These had the radiologic appearance of perforation on cardiac CT; however, they were not associated with pericardial effusions or significant changes in RV pacing lead sensing, impedance, and threshold measurements. CONCLUSIONS: Cardiac CT scanning with multiplanar reformatting is useful for documenting lead position and assessing for possible cardiac perforation. The clinical significance and natural history of leads with only the appearance of perforation on cardiac CT is uncertain.
UR - http://onlinelibrary.wiley.com/doi/10.1111/pace.12307/pdf
UR - https://www.scopus.com/pages/publications/84899648616
U2 - 10.1111/pace.12307
DO - 10.1111/pace.12307
M3 - Article
SN - 0147-8389
VL - 37
SP - 537
EP - 545
JO - Pacing and Clinical Electrophysiology
JF - Pacing and Clinical Electrophysiology
IS - 5
ER -