TY - JOUR
T1 - Method for the calibration of atomic force microscope cantilevers
AU - Sader, John E.
AU - Larson, Ian
AU - Mulvaney, Paul
AU - White, Lee R.
PY - 1995/12/1
Y1 - 1995/12/1
N2 - The determination of the spring constants of atomic force microscope (AFM) cantilevers is of fundamental importance to users of the AFM. In this paper, a fast and nondestructive method for the evaluation of the spring constant which relies solely on the determination of the unloaded resonant frequency of the cantilever, a knowledge of its density or mass, and its dimensions is proposed. This is in contrast to the method of Cleveland et al. [Rev. Sci. Instrum. 64, 403 (1993)], which requires the attachment of masses to the cantilever in the determination of the spring constant. A number of factors which can influence the resonant frequency are examined, in particular (i) gold coating, which can result in a dramatic variation in the resonant frequency, for which a theoretical account is presented and (ii) air damping which, it is found, leads to a shift of ∼4% in the resonant frequency down on its value in a vacuum. Furthermore, the point of load on the cantilever is found to be extremely important, since a small variation in the load point can lead to a dramatic variation in the spring constant. Theoretical results that account for this variation, which, it is believed will be of great practical value to the users of the AFM, are given.
AB - The determination of the spring constants of atomic force microscope (AFM) cantilevers is of fundamental importance to users of the AFM. In this paper, a fast and nondestructive method for the evaluation of the spring constant which relies solely on the determination of the unloaded resonant frequency of the cantilever, a knowledge of its density or mass, and its dimensions is proposed. This is in contrast to the method of Cleveland et al. [Rev. Sci. Instrum. 64, 403 (1993)], which requires the attachment of masses to the cantilever in the determination of the spring constant. A number of factors which can influence the resonant frequency are examined, in particular (i) gold coating, which can result in a dramatic variation in the resonant frequency, for which a theoretical account is presented and (ii) air damping which, it is found, leads to a shift of ∼4% in the resonant frequency down on its value in a vacuum. Furthermore, the point of load on the cantilever is found to be extremely important, since a small variation in the load point can lead to a dramatic variation in the spring constant. Theoretical results that account for this variation, which, it is believed will be of great practical value to the users of the AFM, are given.
UR - http://www.scopus.com/inward/record.url?scp=36449002856&partnerID=8YFLogxK
U2 - 10.1063/1.1145439
DO - 10.1063/1.1145439
M3 - Article
AN - SCOPUS:36449002856
SN - 0034-6748
VL - 66
SP - 3789
EP - 3798
JO - Review of Scientific Instruments
JF - Review of Scientific Instruments
IS - 7
ER -