Abstract
The performance of short drop-time a.c. polarography has been examined in detail, and found to be better than that for natural drop-times in almost every respect. Uncompensated resistance terms are smaller and potentiostat stability is improved. In addition, faster potential scanrates and more rapid data acquisition are possible. A variation of the drop-time over at least three orders of magnitude is possible and this, coupled with excellent instrumental performance, should offer considerable scope in studies of electrode kinetics.
| Original language | English |
|---|---|
| Pages (from-to) | 591-600 |
| Number of pages | 10 |
| Journal | Talanta |
| Volume | 21 |
| Issue number | 6 |
| DOIs | |
| Publication status | Published - Jun 1974 |
| Externally published | Yes |