Abstract
This paper describes a new validated high performance liquid chromatography (HPLC) method for the simultaneous determination of two anti-cancer drugs, Arabinoside-C (Ara-C) and doxorubicin hydrochloride (DOX). A simultaneous determination method saves cost and time as both drugs can be injected into a single HPLC system without the need to change or re-equilibrate with a new mobile phase. The objective of the study is to develop a simultaneous determination method of two anti-cancer drugs, Ara-C and DOX. The mobile phase consisted of a mixture (45:55) of acetonitrile:ammonium hydrogen phosphate aqueous solution (0.01M) at pH 6.2 at a flow rate of 0.3ml/min, with UV detection at 252nm. Separation was achieved on a C-18 column (5 m: 250mm×4.6mm) maintained at 30°C in a column oven. The method was linear between 325ng/ml and 10 g/ml for Ara-C and 625ng/ml and 20 g/ml for DOX. The limit of detection (LOD) was 20ng/ml for Ara-C and 60ng/ml for DOX. The developed HPLC method achieved good precision and accuracy as well as limit of quantitations. The developed and validated method is suitable to be used for routine analysis of Ara-C and DOX.
| Original language | English |
|---|---|
| Pages (from-to) | 472-481 |
| Number of pages | 10 |
| Journal | Toxicology Mechanisms and Methods |
| Volume | 20 |
| Issue number | 8 |
| DOIs | |
| Publication status | Published - Oct 2010 |
| Externally published | Yes |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- Arabinoside-C
- Doxorubicin
- HPLC
- Simultaneous detection
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver