N2 - OBJECTIVES: To establish the success rate, time taken and expertise required to place a feeding catheter into the small intestine using a device that permits real-time localisation of the catheter through detection of an electromagnetic field. DESIGN: Prospective observational study. SETTING: A tertiary, mixed medical-surgical, adult intensive care unit, between February 2008 and February 2009. PARTICIPANTS: 60 attempts at postpyloric intubation by eight clinicians (consultant and trainee intensivists and a dietitian) in 57 critically ill patients who were undergoing mechanical ventilation and were receiving, or suitable to receive, enteral nutrition. MAIN OUTCOME MEASURES: Patients were classified into an initial group (10 patients), who had the catheter placed by a single clinician to establish the most effective technique, and a subsequent group (50 patients), who had the catheter placed by any of the eight clinicians. Catheter position was confirmed on abdominal x-ray by an independent radiologist. RESULTS: Postpyloric catheter placement was successful in 56/60 attempts (93%), in 54/57 patients (95%), with a median time to placement of 7.2 min (interquartile range [IQR], 4.3-12.5 min). In the initial group, placement was successful in 7/10 attempts (70%), with a median time to placement of 20.8 min (IQR, 9.5-32.3 min), compared with 49/50 attempts (98%) and a median time of 5.9 min (IQR, 3.9-11.9 min) in the subsequent group (P = 0.003). CONCLUSIONS: An electromagnetically guided device enabled reliable and rapid bedside placement of small intestinal feeding catheters. Proficiency with the technique was quickly developed and successfully disseminated to a range of clinicians.
AB - OBJECTIVES: To establish the success rate, time taken and expertise required to place a feeding catheter into the small intestine using a device that permits real-time localisation of the catheter through detection of an electromagnetic field. DESIGN: Prospective observational study. SETTING: A tertiary, mixed medical-surgical, adult intensive care unit, between February 2008 and February 2009. PARTICIPANTS: 60 attempts at postpyloric intubation by eight clinicians (consultant and trainee intensivists and a dietitian) in 57 critically ill patients who were undergoing mechanical ventilation and were receiving, or suitable to receive, enteral nutrition. MAIN OUTCOME MEASURES: Patients were classified into an initial group (10 patients), who had the catheter placed by a single clinician to establish the most effective technique, and a subsequent group (50 patients), who had the catheter placed by any of the eight clinicians. Catheter position was confirmed on abdominal x-ray by an independent radiologist. RESULTS: Postpyloric catheter placement was successful in 56/60 attempts (93%), in 54/57 patients (95%), with a median time to placement of 7.2 min (interquartile range [IQR], 4.3-12.5 min). In the initial group, placement was successful in 7/10 attempts (70%), with a median time to placement of 20.8 min (IQR, 9.5-32.3 min), compared with 49/50 attempts (98%) and a median time of 5.9 min (IQR, 3.9-11.9 min) in the subsequent group (P = 0.003). CONCLUSIONS: An electromagnetically guided device enabled reliable and rapid bedside placement of small intestinal feeding catheters. Proficiency with the technique was quickly developed and successfully disseminated to a range of clinicians.
UR - https://www.scopus.com/pages/publications/70350066275
M3 - Article
C2 - 19737119
AN - SCOPUS:70350066275
SN - 1441-2772
VL - 11
SP - 180
EP - 183
JO - Critical Care and Resuscitation
JF - Critical Care and Resuscitation
IS - 3
ER -