Abstract
Dissolution testing for inhalers were previously conducted either on unfractionated drug-carrier powders or drug of specific aerodynamic particle size. In this study, the collection of the full fine particle fraction (FPF) was attempted on a single stage. Capsules containing 30 mg of 2% salbutamol sulfate (SS) was tested to have a FPF of 9 ± 1% using the full set of Andersen cascade impactor (ACI) and a modified Rotahaler® capable of achieving 4.0 kPa pressure drop at 60 L/min air flow rate. A truncated ACI comprising the USP throat, pre-separator, stage 0, stage 4, stage F, polytetrafluoroethylene funnel (TF) and small collection plate (sCP) was found to be capable of achieving a FPF of 9% collected on TF and sCP. An adhesive tape was used to collect the FPF from the TF and sCP and held in place by an enhancer cell in a 200 mL round bottom vessel containing 50 mL Gamble's solution with 0.2 v/v, % Tween 80. Dissolution testing of SS and Seretide® showed burst release of SS and salmeterol while sustained release of fluticasone. This study demonstrated a reproducible method which may be used for evaluation of the full FPF of orally inhaled products.
| Original language | English |
|---|---|
| Pages (from-to) | 45-50 |
| Number of pages | 6 |
| Journal | International Journal of Pharmaceutics |
| Volume | 545 |
| Issue number | 1-2 |
| DOIs | |
| Publication status | Published - 10 Jul 2018 |
| Externally published | Yes |
Keywords
- Andersen cascade impactor
- Dissolution testing
- Enhancer cell
- Fine particle fraction
- Pulmonary delivery
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver