Diethyl ether interference with infrared breath analysis

  • C. M. Bell
  • , S. J. Gutowski
  • , S. Young
  • , D. Wells

Research output: Contribution to journal › Article › Research › peer-review

Abstract

Diethyl ether vapor may substantially Interfere with breath alcohol analysis by instruments based on infrared absorption at 9.5 µm. Exposure of two volunteers simultaneously to diethyl ether vapor for one hour followed immediately by breath tests on the Draeger Alcotest 7110, Siemens Alcomat V5.2F, and Seres Ethylometre 679T produced apparent alcohol readings in one subject of 0.4, 0.1, and 0.1 g/100 mL of blood, respectively. Positive readings persisted in this subject for more than 3 hours. The second subject produced much lower readings of 0.03, 0.01, and 0.00, respectively. Readings persisted with the Alcotest 7110 for one hour. Gas chromatographic analyses of blood and breath samples confirmed that these readings were caused by diethyl ether and not ethanol. The blood concentration of diethyl ether in Subject A immediately after exposure was 25 mg/L. This level produced no clinically detectable neurological changes in the subject.

Original languageEnglish
Pages (from-to)166-168
Number of pages3
JournalJournal of Analytical Toxicology
Volume16
Issue number3
DOIs
Publication statusPublished - May 1992
Externally publishedYes