Developing small vacuum spark as an x-ray source for calibration of an x-ray focusing crystal spectrometer

Mostafa Ghomeishi, Mohammad Karami, Faisal Rafiq Mahamd Adikan

Research output: Contribution to journalArticleResearchpeer-review

Abstract

A new technique of x-ray focusing crystal spectrometers' calibration is the desired result. For this purpose the spectrometer is designed to register radiated copper Kα and Kβ lines by using a flat α-quartz crystal. This experiment uses pre-breakdown x-ray emissions in low vacuum of about 2.5-3 mbar. At this pressure the pinch will not form so the plasma will not radiate. The anode material is copper and the capacity of the capacitor bank is 22.6 nF. This experiment designed and mounted a repetitive triggering system to save the operator time making hundreds of shots. This emission amount is good for calibration and geometrical adjustment of an optical crystal x-ray focusing spectrometer.

Original languageEnglish
Article number103110
JournalReview of Scientific Instruments
Volume83
Issue number10
DOIs
Publication statusPublished - 29 Oct 2012
Externally publishedYes