TY - JOUR
T1 - Delivery of positive end-expiratory pressure to preterm lambs using common resuscitation devices
AU - Thio, Marta
AU - Dawson, Jennifer A.
AU - Crossley, Kelly J.
AU - Moss, Timothy J.
AU - Roehr, Charles C.
AU - Polglase, Graeme R.
AU - Davis, Peter G.
AU - Hooper, Stuart B.
PY - 2019/1/1
Y1 - 2019/1/1
N2 - Background: In neonatal resuscitation, a ventilation device providing positive end-expiratory pressure (PEEP) is recommended. There is limited information about PEEP delivery in vivo, using different models of self-inflating bag (SIB) at different inflation rates and PEEP settings. Methods: We compared PEEP delivery to intubated preterm lambs using four commonly available models of paired SIBs and PEEP valves, with a T-piece, with gas flow of 8 L/min. Peak inspiratory pressure inflations of 30 cmH2O, combined with set PEEP of 5, 7 and 10 cmH2O, were delivered at rates of 20, 40 and 60/min. These combinations were repeated without gas flow. We measured mean PEEP, maximum and minimum PEEP, and its difference (PEEP reduction). Results: A total of 3288 inflations were analysed. The mean PEEP delivered by all SIBs was lower than set PEEP (P<0.001), although some differences were <0.5 cmH2O. In 55% of combinations, the presence of gas flow resulted in increased PEEP delivery (range difference 0.3-2 cmH2O). The mean PEEP was closer to set PEEP with faster inflation rates and higher set PEEPs. The mean (SD) PEEP reduction was 3.9 (1.6), 8.2 (1.8), 2 (0.6) and 1.1 (0.6) cmH2O with the four SIBs, whereas it was 0.5 (0.2) cmH2O with the T-piece. Conclusions: PEEP delivery with SIBs depends on the set PEEP, inflation rate, device model and gas flow. At recommended inflation rates of 60/min, some devices can deliver PEEP close to the set level, although the reduction in PEEP makes some SIBs potentially less effective for lung recruitment than a T-piece.
AB - Background: In neonatal resuscitation, a ventilation device providing positive end-expiratory pressure (PEEP) is recommended. There is limited information about PEEP delivery in vivo, using different models of self-inflating bag (SIB) at different inflation rates and PEEP settings. Methods: We compared PEEP delivery to intubated preterm lambs using four commonly available models of paired SIBs and PEEP valves, with a T-piece, with gas flow of 8 L/min. Peak inspiratory pressure inflations of 30 cmH2O, combined with set PEEP of 5, 7 and 10 cmH2O, were delivered at rates of 20, 40 and 60/min. These combinations were repeated without gas flow. We measured mean PEEP, maximum and minimum PEEP, and its difference (PEEP reduction). Results: A total of 3288 inflations were analysed. The mean PEEP delivered by all SIBs was lower than set PEEP (P<0.001), although some differences were <0.5 cmH2O. In 55% of combinations, the presence of gas flow resulted in increased PEEP delivery (range difference 0.3-2 cmH2O). The mean PEEP was closer to set PEEP with faster inflation rates and higher set PEEPs. The mean (SD) PEEP reduction was 3.9 (1.6), 8.2 (1.8), 2 (0.6) and 1.1 (0.6) cmH2O with the four SIBs, whereas it was 0.5 (0.2) cmH2O with the T-piece. Conclusions: PEEP delivery with SIBs depends on the set PEEP, inflation rate, device model and gas flow. At recommended inflation rates of 60/min, some devices can deliver PEEP close to the set level, although the reduction in PEEP makes some SIBs potentially less effective for lung recruitment than a T-piece.
KW - low resource setting
KW - neonatal resuscitation
KW - PEEP delivery
KW - self-inflating bag
KW - T-piece
UR - http://www.scopus.com/inward/record.url?scp=85049007410&partnerID=8YFLogxK
U2 - 10.1136/archdischild-2017-314064
DO - 10.1136/archdischild-2017-314064
M3 - Article
AN - SCOPUS:85049007410
VL - 104
SP - F83-F88
JO - Archives of Disease in Childhood: Fetal and Neonatal Edition
JF - Archives of Disease in Childhood: Fetal and Neonatal Edition
SN - 1359-2998
IS - 1
ER -