Continuous measurement of arterial and end-tidal carbon dioxide during cardiac surgery: P(a-ET)CO2 gradient

Paul S. Myles, D. A. Story, M. A. Higgs, M. R. Buckland

Research output: Contribution to journalArticleResearchpeer-review

11 Citations (Scopus)

Abstract

There have been reports of a negative arterial to end-tidal CO2 gradient (P(a-ET)CO2) during cardiac surgery, so we used capnometry and an intravascular blood gas sensor (Paratrend 7) to continuously monitor this gradient in 20 cardiac surgical patients. We also compared the values obtained from this sensor with those obtained from a standard blood gas analyser at seven time points. We found a significant change in P(a-ET)CO2 after cardopulmonary bypass (P < 0.001) though we were unable to demonstrate a negative P(a-ET)CO2 at any time (95% CI 0.14%). There was clinically acceptable agreement between labarotary and Paratrend 7 measurements during and after cardiac surgery.

Original languageEnglish
Pages (from-to)459-463
Number of pages5
JournalAnaesthesia and Intensive Care
Volume25
Issue number5
Publication statusPublished - 1997

Keywords

  • Anaesthesia: monitoring, blood gas sensors, capnography, cardiac surgery