TY - JOUR
T1 - Comparison of CT on Rails With Electronic Portal Imaging for Positioning of Prostate Cancer Patients With Implanted Fiducial Markers
AU - Owen, Rebecca
AU - Kron, Tomas
AU - Foroudi, Farshad
AU - Milner, Alvin
AU - Cox, Jennifer
AU - Duchesne, Gillian
AU - Cleeve, Laurence
AU - Zhu, Li
AU - Cramb, Jim
AU - Sparks, Laura
AU - Laferlita, Marcus
PY - 2009/7/1
Y1 - 2009/7/1
N2 - Purpose: The objective of this investigation was to measure the agreement between in-room computed tomography (CT) on rails and electronic portal image (EPI) radiography. Methods and Materials: Agreement between the location of the center of gravity (COG) of fiducial markers (FMs) on CT and EPI images was determined in phantom studies and a patient cohort. A secondary analysis between the center of volume (COV) of the prostate on CT and the COG of FMs on CT and EPI was performed. Agreement was defined as the 95% probability of a difference of ≤3.0 mm between images. Systematic and random errors from CT and EPI are reported. Results: From 8 patients, 254 CT and EPI pairs were analyzed. FMs were localized to within 3 mm on CT and EPI images 96.9% of the time in the left-right (LR) plane, 85.8% superior-inferior (SI), and 89% anterior-posterior (AP). The differences between the COV on CT and the COG on EPI were not within 3 mm in any plane: 87.8% (LR), 64.2% (SI), and 70.9% (AP). The systematic error varied from 1.2 to 2.9 mm (SI) and 1.8-2.9 mm (AP) between the COG on EPI and COV on CT. Conclusions: Considerable differences between in-room CT and EPI exist. The phantom measurements showed slice thickness affected the accuracy of localization in the SI plane, and couch sag that occurs at the CT on rails gantry could not be totally corrected for in the AP plane. Other confounding factors are the action of rotating the couch and associated time lag between image acquisitions (prostate motion), EPI image quality, and outlining uncertainties. Crown
AB - Purpose: The objective of this investigation was to measure the agreement between in-room computed tomography (CT) on rails and electronic portal image (EPI) radiography. Methods and Materials: Agreement between the location of the center of gravity (COG) of fiducial markers (FMs) on CT and EPI images was determined in phantom studies and a patient cohort. A secondary analysis between the center of volume (COV) of the prostate on CT and the COG of FMs on CT and EPI was performed. Agreement was defined as the 95% probability of a difference of ≤3.0 mm between images. Systematic and random errors from CT and EPI are reported. Results: From 8 patients, 254 CT and EPI pairs were analyzed. FMs were localized to within 3 mm on CT and EPI images 96.9% of the time in the left-right (LR) plane, 85.8% superior-inferior (SI), and 89% anterior-posterior (AP). The differences between the COV on CT and the COG on EPI were not within 3 mm in any plane: 87.8% (LR), 64.2% (SI), and 70.9% (AP). The systematic error varied from 1.2 to 2.9 mm (SI) and 1.8-2.9 mm (AP) between the COG on EPI and COV on CT. Conclusions: Considerable differences between in-room CT and EPI exist. The phantom measurements showed slice thickness affected the accuracy of localization in the SI plane, and couch sag that occurs at the CT on rails gantry could not be totally corrected for in the AP plane. Other confounding factors are the action of rotating the couch and associated time lag between image acquisitions (prostate motion), EPI image quality, and outlining uncertainties. Crown
KW - EPI
KW - FM
KW - IGRT
KW - In-room CT
UR - http://www.scopus.com/inward/record.url?scp=67049086550&partnerID=8YFLogxK
U2 - 10.1016/j.ijrobp.2009.01.054
DO - 10.1016/j.ijrobp.2009.01.054
M3 - Article
C2 - 19480970
AN - SCOPUS:67049086550
SN - 0360-3016
VL - 74
SP - 906
EP - 912
JO - International Journal of Radiation Oncology Biology Physics
JF - International Journal of Radiation Oncology Biology Physics
IS - 3
ER -