Abstract
Sixty-one consecutive patients in the Intensive Care Unit requiring central venous lines (CVC) for five or more days were randomized to receive either a standard triple lumen CVC (STD/CVC) or a silver sulphadiazine and chlorhexidine impregnated CVC (SSD/CVC). Data from the 54 patients who completed the trial show a reduced infection rate (positive tip culture) in the SSD/CVC group (4 out of 28) compared to the STD/CVC group (10 out of 26) (P < 0.05). In addition, the new Fibrin Analysing System (FAS) brush was evaluated and used to determine die presence of infection in all the CVCs (STD/CVC and SSD/CVC combined, n = 54) at day 3 (i.e. early warning of CVC colonization/infection) and at the time of removal of the CVC. The FAS brush was able to detect an infected CVC on only one occasion on day 3 out of the 14 CVC tips which were later found to be colonized/infected at the time of removal. The sensitivity of the FAS brush in detecting colonized/infected CVCs at the time of CVC removal compared with CVC tip culture was 21% with a specificity of 100%. These findings would currently not support the routine use of the FAS blush in determining CVC infection/colonization.
Original language | English |
---|---|
Pages (from-to) | 330-333 |
Number of pages | 4 |
Journal | Anaesthesia and Intensive Care |
Volume | 24 |
Issue number | 3 |
DOIs | |
Publication status | Published - Jun 1996 |
Externally published | Yes |
Keywords
- Equipment, infection: venous catheters, central