7-Tesla MRI of the brain in a research subject with bilateral, total knee replacement implants: Case report and proposed safety guidelines

Giuseppe Barisano, Bozena Culo, Frank G. Shellock, Farshid Sepehrband, Katherin Martin, Mary Stevens, Danny J. Wang, Arthur W. Toga, Meng Law

Research output: Contribution to journalArticleOtherpeer-review

Abstract

Recently, the first 7-T MR system was approved for clinical use in the United States. Unfortunately, relatively few metallic implants have undergone testing to determine if they are acceptable or pose hazards to research subjects and patients at this ultra-high-field strength. Therefore, in lieu of not performing a research or clinical MRI exam at 7-T, the supervising physician may make a decision to scan the individual with an untested metallic implant based on an analysis of the risks vs. the benefits. We present a case report of a research subject with bilateral, total knee replacement implants that safely underwent MRI of the brain at 7-T and provide guidelines for healthcare professionals to follow in order to ensure safety in research subjects or patients with metallic implants referred for 7-T scans.

Original languageEnglish
Pages (from-to)313-316
Number of pages4
JournalMagnetic Resonance Imaging
Volume57
DOIs
Publication statusPublished - 1 Apr 2019
Externally publishedYes

Keywords

  • 7-Tesla
  • Brain
  • Magnetic resonance imaging (MRI)
  • MRI implants
  • MRI safety

Cite this

Barisano, Giuseppe ; Culo, Bozena ; Shellock, Frank G. ; Sepehrband, Farshid ; Martin, Katherin ; Stevens, Mary ; Wang, Danny J. ; Toga, Arthur W. ; Law, Meng. / 7-Tesla MRI of the brain in a research subject with bilateral, total knee replacement implants : Case report and proposed safety guidelines. In: Magnetic Resonance Imaging. 2019 ; Vol. 57. pp. 313-316.
@article{434367533085474fad463e181319d8ec,
title = "7-Tesla MRI of the brain in a research subject with bilateral, total knee replacement implants: Case report and proposed safety guidelines",
abstract = "Recently, the first 7-T MR system was approved for clinical use in the United States. Unfortunately, relatively few metallic implants have undergone testing to determine if they are acceptable or pose hazards to research subjects and patients at this ultra-high-field strength. Therefore, in lieu of not performing a research or clinical MRI exam at 7-T, the supervising physician may make a decision to scan the individual with an untested metallic implant based on an analysis of the risks vs. the benefits. We present a case report of a research subject with bilateral, total knee replacement implants that safely underwent MRI of the brain at 7-T and provide guidelines for healthcare professionals to follow in order to ensure safety in research subjects or patients with metallic implants referred for 7-T scans.",
keywords = "7-Tesla, Brain, Magnetic resonance imaging (MRI), MRI implants, MRI safety",
author = "Giuseppe Barisano and Bozena Culo and Shellock, {Frank G.} and Farshid Sepehrband and Katherin Martin and Mary Stevens and Wang, {Danny J.} and Toga, {Arthur W.} and Meng Law",
year = "2019",
month = "4",
day = "1",
doi = "10.1016/j.mri.2018.11.016",
language = "English",
volume = "57",
pages = "313--316",
journal = "Magnetic Resonance Imaging",
issn = "0730-725X",
publisher = "Elsevier",

}

7-Tesla MRI of the brain in a research subject with bilateral, total knee replacement implants : Case report and proposed safety guidelines. / Barisano, Giuseppe; Culo, Bozena; Shellock, Frank G.; Sepehrband, Farshid; Martin, Katherin; Stevens, Mary; Wang, Danny J.; Toga, Arthur W.; Law, Meng.

In: Magnetic Resonance Imaging, Vol. 57, 01.04.2019, p. 313-316.

Research output: Contribution to journalArticleOtherpeer-review

TY - JOUR

T1 - 7-Tesla MRI of the brain in a research subject with bilateral, total knee replacement implants

T2 - Case report and proposed safety guidelines

AU - Barisano, Giuseppe

AU - Culo, Bozena

AU - Shellock, Frank G.

AU - Sepehrband, Farshid

AU - Martin, Katherin

AU - Stevens, Mary

AU - Wang, Danny J.

AU - Toga, Arthur W.

AU - Law, Meng

PY - 2019/4/1

Y1 - 2019/4/1

N2 - Recently, the first 7-T MR system was approved for clinical use in the United States. Unfortunately, relatively few metallic implants have undergone testing to determine if they are acceptable or pose hazards to research subjects and patients at this ultra-high-field strength. Therefore, in lieu of not performing a research or clinical MRI exam at 7-T, the supervising physician may make a decision to scan the individual with an untested metallic implant based on an analysis of the risks vs. the benefits. We present a case report of a research subject with bilateral, total knee replacement implants that safely underwent MRI of the brain at 7-T and provide guidelines for healthcare professionals to follow in order to ensure safety in research subjects or patients with metallic implants referred for 7-T scans.

AB - Recently, the first 7-T MR system was approved for clinical use in the United States. Unfortunately, relatively few metallic implants have undergone testing to determine if they are acceptable or pose hazards to research subjects and patients at this ultra-high-field strength. Therefore, in lieu of not performing a research or clinical MRI exam at 7-T, the supervising physician may make a decision to scan the individual with an untested metallic implant based on an analysis of the risks vs. the benefits. We present a case report of a research subject with bilateral, total knee replacement implants that safely underwent MRI of the brain at 7-T and provide guidelines for healthcare professionals to follow in order to ensure safety in research subjects or patients with metallic implants referred for 7-T scans.

KW - 7-Tesla

KW - Brain

KW - Magnetic resonance imaging (MRI)

KW - MRI implants

KW - MRI safety

UR - http://www.scopus.com/inward/record.url?scp=85059139992&partnerID=8YFLogxK

U2 - 10.1016/j.mri.2018.11.016

DO - 10.1016/j.mri.2018.11.016

M3 - Article

VL - 57

SP - 313

EP - 316

JO - Magnetic Resonance Imaging

JF - Magnetic Resonance Imaging

SN - 0730-725X

ER -